Model FL-PWJ Power Supply Model FL-LHJ Lamphouse Instruction Manual
The system should be operated in a room with as little dust as practically possible. Keep the power supply and lamphouse away from solvents, chemical
2.3 Set Up • As a first step, remove all components from the shipping container and remove the packing materials. Save the container and pack
• Mount the lamp house onto the microscope. The three screws that engage the dovetail fitting should be equally tight, the lamp house should be mount
2.4 Bulb Installation It is critical to properly install the bulb in the socket. Reversed polarity can dramatically shorten the life expecta
3. Slide desired lamp into the lamp holder. 4. Slide the bulb clamp onto the negative (-) end of the lamp. 5. Hold the bulb clamp and trigger lead
2.5 Bulb Alignment – TC Models After the bulb has burned in, the system is ready for arc lamp alignment. There are many methods to accompl
similar. 3) Open the filter slider on the illuminator so that the light travels through the system. Engage the field iris diaphragm all the way ope
7) Check the effect achieved when the collector lens focus knob is adjusted and continue to make minor adjustments to the arc position and the mirror
similar. 3) Open the filter slider on the illuminator so that the light travels through the system. Engage the field iris diaphragm all the way ope
7) Check the effect achieved when the collector lens focus knob is adjusted and continue to make minor adjustments to the arc position and the mirror
Table of Contents 1.0 Introduction 1.1 Power Supply & Lamphouse Features 1.2 General Safety Guidelines 1.3 Warning/Caution symbols used in
Always disconnect the power cord from the back of the power supply when not being used, when cleaning your instrument or when making any repairs.
UV light causes sunburn. Long term exposure can result in loss of skin elasticity initially and carcinoma eventually. Absorption of UV light by the e
• In warm humid climates, take special care to prevent your equipment from exposure to fungal growth. Dust Protection Be sure to use the supplied
Meiji Techno products are manufactured in Japan under ISO9000 manufacturing standards. However, if you ever have any difficulty with any Meiji product
Refer to the companion manual of the Nobska Lamphouse for more on troubleshooting and replacement of the 100W Mercury Lamp. Replacing the mercury HB
Input 90 – 132 VAC 50-60Hz Output 20VDC @ 5A Output Stability (Ripple) < 2.5% P-P Starting Voltage Peak 2.3KV DC Operating Temperature 0 – 40
Meiji Techno Co. Ltd. MEIJI TECHNO AMERICA 3010 Olcott St Santa Clara, CA 95054 Toll free: Phone: Fax: Email: Web: (800) 832-0060
Model FL-PWJ Power Supply & FL-LHJ Lamphouse 1. Introduction The Meiji Techno FL-PWJ Epi-Fluorescence Illuminator Power Supply is a precision
1.1 Power Supply & Lamphouse Features • Integrated Focusing Collector Lens in Lamphouse • Integrated Concave Mirror optimizes DC Arc Component
These symbols are used throughout this instruction manual. For your safety, be sure to follow all instructions associated with the symbols listed belo
Disassembly of the product may result in electrical shock, injury or death, equipment damage, loss of warranty coverage or may create other potential
furnished under this warranty are subject to the same terms and conditions of the original warranty. This warranty supersedes any other warranty and i
time. In some cases, requests for warranty service are received which are not justified. In these cases, Meiji Techno will provide an explanation for
1. Power Switch 2. Run Indicator Lamp 3. Panel Mounted Elapsed Time Meter 4. Reset Button The image above indicates the main components of the ba